For care of the material, please do not fail to comply with the following important points:
If sterilized items are not used immediately, the sterile packaging must at least be
labeled with the expiration date of sterility and additionally a sterile indicator should
be added.
6.1 Control unit and foot pedal
The patient should not come into contact with the control unit and the foot pedal:
Only carry out outside wiping disinfection with microbiologically tested surface disinfectants or 80% vol.
ethylic alcohol.
Only wipe control unit with cloth. Please do not perform spray cleaning or spray
disinfection, because the unit is not sealed!
6.2 Electronic motor 31ESS
Restrictions of treatment Frequent reprocessing hardly affects the electronic motor. The end of
INSTRUCTIONS
Usage location:
Storage and transport:
Cleaning preparations:
Automatic cleaning and
disinfecting:
8
6 Cleaning and disinfection
product life is usually determined by damage, wear and tear caused by
usage.
Wipe dirty surfaces with an expendable cloth/ tissue paper.
No special requirements. Avoid lengthy waiting periods before preparation
because of danger of drying and corrosion.
Remove any surface soiling of the electronic motor with a disposable
towel. Unscrew motor cover and remove it with the cable.
Equipment: Cleaning/disinfection equipment with special load carrier,
which ensures the connection of the motors to the cleaning/disinfecting
equipment and the irrigation of the channels. Wash the electronic motor
from the front side. Use neutral cleaners as cleansing agents.
1) Load the electronic motor in the load carrier (irrigation of the channels
must be ensured).
2) Load Motor cover with cable in basket.
3) Set cleaning cycle with effectual rinsing and cleaning. Carry out rinse
cycle with demineralized water.
4) Rinse with demineralized water at 93°C for 10 minutes for thermal
disinfecting.
5) Check whether there is still visible dirt in the gaps when taking the
electronic motor and motor cover with cable out. If necessary repeat
cycle or clean manually.
No. 31739 26/12