Step-by-Step Instructions
6.6.5
Completion of the measurements
Before unplugging the measurement cable from the device
receptacle disconnect the lip clip and the file clip from the patient.
Move the file stopper to the selected reference point on the tooth.
Gently remove the file from the canal and measure the apical length
between the stopper and the file.
Note
As a safety precaution in order to avoid over-instrumentation, it is
recommended to proceed as follows: stop file progression in the
canal at the point where the Propex Pixi
'0.0'.
Place the file onto an endodontic ruler and measure the apical length.
Subtract a minimum of 0.5 mm from the measured file length.
6.7
Sound Adjustment
Propex Pixi
enables monitoring of the progression of the file within the canal in
addition to visual monitoring.
The volume can be adjusted to four different levels: mute, low, normal
and high, by successively pressing the
When sound level is muted the
levels the
22/1346
®
Apex Locator is equipped with an audio indicator which
icon remains turned on.
®
Apex Locator indicates
"VOLUME" button.
icon is turned off. At other sound

Publicidad