Maintenance, Cleaning, Disinfection and Sterilization
7
Maintenance, Cleaning, Disinfection and
Sterilization
7.1
General Recommendations
The device does not contain user serviceable parts. The service
and repair should be provided by factory trained service personnel
only.
After each use, all objects that were in contact with infectious
agents should be cleaned using single-use wipes or a soft cloth
impregnated with a disinfecting and detergent solution (a
bactericidal, fungicidal and aldehyde free solution), according to
manufacturer's directions. We recommend using only a disinfecting
solution which is approved for its efficacy (VAH/DGHM-listing, CE
marking, FDA approval). Use of chemical agents may cause
damage to the equipment. To clean the device housing and
measuring cable, we recommend single-use wipes (e.g.
DENTIRO
be removed. No visible impurities or liquid should remain on the
device housing after disinfection.
In addition, the lip clip and the file clip must be sterilized between
treatments. Please note that the device housing, the AC adapter
and the measuring cable cannot be sterilized.
In addition, a fork is not included in the Propex Pixi
but it can be used and should follow the same sterilization
procedure as the lip clip and file clip.
Follow the "Disinfection and sterilization procedure" described in
section
The user is responsible for the sterility of the lip clip, the file clip and
the fork for the first cycle and each further usage.
All damaged accessories should be discarded, while dirty
accessories should be cleaned and sterilized per the procedure
described in section 7.2.
®
Wipes). After cleaning, all disinfectant residues should
7.2
for the lip clip, the file clip and the fork.
®
Apex Locator
25/1346