Maintenance, Cleaning, Disinfection and Sterilization
Operation
9
Sterilization
10
Storage
32/1346
Operating mode
Steam sterilization at:
134 °C / 273°F,
3 min.
Keep devices in
sterilization
packaging in a dry
and clean
environment.
Warning
The accessories (lip clip, file
clip and fork) must be
sterilized according to the
packaging labelling.
Use only autoclaves that
match the requirements of
EN 13060, EN 285.
Respect the maintenance
procedure of the autoclave
device given by the
manufacturer.
Use only this recommended
sterilization procedure.
Control the efficiency
(packaging integrity, no
humidity, colour change of
sterilization indicators,
physico-chemical
integrators, digital records of
cycles parameters).
Sterility cannot be
guaranteed if packaging is
open, damaged or wet.
Check the packaging and the
medical devices before using
them (packaging integrity, no
humidity and validity period).