Troubleshooting
#
Problem
The device does not turn
2
on by pressing the "ON /
OFF" button.
The device shuts off during
3
the procedure.
No sound during the
4
procedure.
Display is not steady
5
during the procedure.
Possible cause
The button is
functioning
incorrectly.
The battery is
discharged.
Electronic
malfunction.
The battery is low.
The sound control
is set at "Mute"
level.
There is not a good
contact between the
lip clip and the oral
mucosa.
The file clip is
soiled.
Deep caries
provides a
conductive path
outside the canal.
Perforation.
Large lateral canal.
Solution
Try pressing the "ON/ OFF"
button several times. If the
device still does not turn
on, please contact your
distributor.
Charge the battery.
Contact your distributor.
Charge the battery.
Adjust the sound level by
pressing the "VOLUME"
button.
Ensure a good contact
between the mucosa and
lip clip (Place the lip clip in
the labial angle opposite
the tooth to be treated).
Clean the file clip (with
Ethanol).
Block the external
conductive path.
Remove the file, close the
perforation and repeat the
apex detection procedure,
carefully inserting the file
into canal.
Try continuing the
procedure by gently
advancing the file.
35/1346