Descargar Imprimir esta página

Dentsply Sirona Propex Pixi Manual De Usuario página 34

Ocultar thumbs Ver también para Propex Pixi:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 197
Troubleshooting
10
Troubleshooting
Please review the checklist below should you experience a problem
with your Propex Pixi
following the proposed solutions, please contact your distributor.
Warning
The following patient related factors may prevent accurate readings:
- Blocked root canals;
- Teeth with large apices;
- Root fracture or perforation;
- Metal crowns or bridges, if they come into contact with the file or
the lip clip.
#
Problem
During battery charging
the battery symbol blinks
at a fast rate.
1
34/1346
®
Apex Locator. If the problem persists after
Possible cause
The battery is not
connected.
The battery is not
rechargeable.
Solution
Open the battery
compartment, and connect
the battery as described in
the User Manual, section
6.4.
Replace the battery with a
rechargeable type as
described in the user
manual, section 6.4.

Publicidad

loading