MAINTENANCE AND USE MANUAL
EN
3.3.3 Indicators icons
*
This icon is displayed shortly after a cycle has started and indicates that the machine
door is locked.
This icon indicates that a cycle is running: you will see the arrows spinning around.
Icon n° 3 warns that printer is out of paper, or the that printer cover is not
properly closed. The machine can execute its cycle without any risk.
Kind of cycle selected.
USB icon: the autoclaves are provided with operating software able to communicate
with and write to a USB flash drive. Some sterilisers incorporate a USB port, while
others can write to a USB flash drive through the special STS Datalogger system
connected to the RS232 serial port.
The icon is displayed when the USB flash drive is inserted in the connector (or the
Datalogger). It is important to read the attached instructions when using either the
USB incorporated in the autoclave or the Datalogger system.
DATA IS SAVED ONLY AND EXCLUSIVELY IF THE FLASH DRIVE HAS ALREADY BEEN
INSERTED IN THE CONNECTOR BEFORE LAUNCHING THE CYCLE.
This icon warns the user that clean water is at its minimum level: it is not possible to
start a new cycle. Fill the machine with water before starting other cycles.
The clean water tank is full. If the PUMP button is pressed again, the display will show
"FULL WATER RESERVOIR", warning the user that it is not possible to continue filling the
tank with water.
While this icon is displayed, no cycle can start: the dirty water tank should be comple-
tely drained. Allow all water to drain out before closing the draining tap (see paragraph
9.2).
* see paragraph 3.3.2.