MANUAL WATER DISCHARGE
9.1
8
Figure
9
Figure
10
Figure
11
Figure
Bacterial filter: replace after approx. 300 cycles or every three months at the latest.
164 of 228
CHAPTER 9: WATER DISCHARGE
Fig. 8: Maximum dirty water level symbol.
Fig. 9: Press DOOR to open.
Fig. 10: Connect the hose (supplied) to the discharge
connector . Push the connector towards the autoclave
and discharge all the water.
Fig. 11:
the task has been completed.
MAINTENANCE AND USE MANUAL
EN
The capacity of the dirty water tank is
4.5 litres.
Dirty water may contain contaminated
residues
and
recommended that protective gloves
are used while draining water from the
equipment.
Wait for all the water to go out
before disconnecting the tube.
Pull the connector and remove the hose once
it
is
therefore
zetaclave B