WARNING
Wear protective equipment according to the tool
manufacturer's instructions
Drill holes and blow out with bellows
Check the distances between the drilled holes
Hold the ceiling bracket to the ceiling and hammer in the
fastening anchor.
Tighten the fastener according to the manufacturer's
instructions.
DANGER
Danger to life through an electric shock.
Do not switch on the mains power supply before the
light head has been installed
To avoid the risk of an electric shock, this device
may only be connected to a supply network with
a protective earth conductor
Create a network connection.
30
3.4
Ceiling bracket installation
A
Pull the cable of the ceiling rod through the opening "B"
of the ceilingbracket.
Insert the ceiling rod "A" in the ceiling bracket.
Secure with locking screw "C" and M8 nut.
D
Tighten locking screw "C" M8 and nut (20 NM).
Tighten all 4 setscrews "D" (5 NM).
Connect the connector of the ceiling rod to the connector
of the power unit.
Slide the ceiling hood and ring over the ceiling bracket
and screw tight.
B
C
C