Descargar Imprimir esta página

Derungs Dmed 60 C Triango Instrucciones De Montaje Y Funcionamiento página 32

Luz de tratamiento

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 88
ENG
4.
INSTALLATION: Triango 60 W
4.1
Load data
Bending moment M
B
Gravitational force F
G
4.2
Wall bracket installation
Fastening material is not included in the scope
of delivery.
DANGER
Installation by qualified personnel
Installation may only be carried out by a qualified
specialist. Without the appropriate knowledge, life-
threatening dangers may present themselves.
Two people are required for the installation
CAUTION
Use the load data table to determine the fasteners
Note the dimensions of the linkage before installation
CAUTION
Note the position of the wall bracket
The wall bracket must be aligned according to the axis
in the picture
Disregarding the correct alignment will result in
mechanical safety defects
For lightweight walls, we recommend a counter plate
(not included in the scope of delivery)
DANGER
Danger to life through an electric shock.
The mains cable must be disconnected from the
mains at all poles by an external lockable switch.
Draw in 6 drilling marks.
Note the position of the opening "A" for the purpose
of power connection.
32
275 Nm
135 N
A
WARNING
Wear protective equipment according to the tool
manufacturer's instructions
Drill holes and blow out with bellows.
Check the distances between the drilled holes.
Position the wall bracket on the wall and hammer in the
fastening anchor.
Tighten the fastener according to the manufacturer's
instructions.
DANGER
Danger to life through an electric shock.
Do not switch on the mains power supply before the
light head has been installed.
To avoid the risk of an electric shock, this device may
only be connected to a supply network with
a protective earth conductor
Create a network connection.