ENG
13. TROUBLESHOOTING
Fault
Light does not come on
Light does not come on
Light does not come on
14. TECHNICAL DATA
Electrical values:
Rated connection voltage
Frequency range
Power consumption:
Triango 60-1 / Triango 60-3
Triango Fokus 60-1
Integrated electronic transformer
Photometric values*:
Central illuminance Ec at 1.0m distance
Illuminated field diameter d10 at a distance of 1.0m:
Triango 60-1 / Triango 60-3
Triango Fokus 60-1
Illuminated field diameter d50 at a distance of 1.0m
Colour temperature:
Triango 60-1 / Triango Fokus 60-1
Triango 60-3
Colour rendering index Ra
Residual illuminance when shaded by a shadow
Residual illuminance when shaded by two shadows
Residual illuminance in tube
Residual illuminance in tube with a shadow
Residual illuminance in tube with two shadows
Illumination depth L1 + L2
40
Possible cause
Contact fault
Light defective
No mains voltage
Troubleshooting
Switch on again
Contact manufacturer
Check voltage, inspect all connections
User profile
All
Only via manufacturer
Electrician
100-240 V
50-60 Hz
21 W (max. 45VA)
33 W (max. 64VA)
24VDC output
60,000 lx
Ø = 18 cm
Ø = 18 cm / 23 cm / 25 cm
Ø = 10 cm
4700K
3700K / 4300K / 4700K
95
< 1%
64.9%
100%
< 1%
65.1%
116 cm
* -10% / +20% tolerance