The above listed models are intended for use in the electromagnetic environment specified below. The customer or the user of
the above listed models should assure that they are used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Radiated RF EM fields
and Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
RATED power
frequency
magnetic fields
IEC 61000-4-8
IMMUNITY to
proximity magnetic
fields
44
Test Condition
±,8 kV Contact
± 2,4,8,15 kV Air
10 V/m
80 MHz – 2,7 GHz
80% AM 1kHz
385MHz (18Hz Pulse Modulation)
450MHz (FM+/-5KHz deviation 1kHz
sine or 18Hz Pulse Modulation)
710MHz (217Hz PM)
745MHz (217Hz PM)
780MHz (217Hz PM)
810MHz (18Hz PM)
870MHz (18Hz PM)
930MHz (18Hz PM)
1720MHz (217Hz PM)
1845MHz (217Hz PM)
1970MHz (217Hz PM)
2450MHz (217Hz PM)
5240MHz (217Hz PM)
5500MHz (217Hz PM)
5785MHz (217Hz PM)
50Hz or 60Hz
30 kHz CW
Test level: 8 A/m
134,2 kHz PM 2,1 kHz
Test level: 65 A/m
13,56 MHz PM 50 kHz
Test level: 7,5 A/m
Enclosure Port
IEC 60601
Compliance level
±,8 kV Contact
± 2,4,8,15 kV Air
10 V/m
80 MHz – 2,7 GHz
27 V/m
28 V/m
9 V/m
9 V/m
9 V/m
28 V/m
28 V/m
28 V/m
28 V/m
28 V/m
28 V/m
28 V/m
9 V/m
9 V/m
9 V/m
30 A/m
Not applicable.
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a professional healthcare facility
environment and Home healthcare
environment.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital environment.
The Triango 60 light do not contain
magnetically sensitive components
or circuitry within the ENCLOSURE.