Descargar Imprimir esta página

Microlife OXY200 Manual Del Usuario página 71

Ocultar thumbs Ver también para OXY200:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 9
EMC Information-Guidance and Manufacturer's Declaration
1* WARNING:
Use of this equipment adjacent to or stacked with other
equipment should be avoided because it may result in improper
operation. If such use is necessary, this equipment and the
other equipment should be observed to verify that they are
operating normally.
2* WARNING:
Use of accessories, transducers and cables other than those
specified or provided by the manufacturer of this equipment
could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this
equipment and may result in improper operation.
3* WARNING:
Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the ME
equipment, including cables specified by the manufacturer.
Otherwise, it may result in the degradation of the performance
of this equipment could result.
Declaration – Electromagnetic
Immunity
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Not applicable
IEC 61000-3-3
Declaration – Electromagnetic Immunity
Immunity test
IEC 60601 test level
±8 kV contact
Electrostatic discharge (ESD)
±2 kV, ±4 kV, ±8 kV,
IEC 61000-4-2
±15 kV air
± 2 kV for power
Electrical fast transient / burst
supply lines
IEC 61000-4-4
± 1 kV for input /
output lines
± 0.5kV, ± 1 kV line(s)
to lines
Surge IEC 61000-4-5
± 0.5kV, ± 1 kV, ± 2 kV
line(s) to earth
0% UT; 0.5 cycle At 0°,
45°, 90°, 135°, 180°,
Voltage dips, short
225°, 270°and 315°
interruptions and voltage
0% UT; 1 cycle and 70%
variations on power supply
UT; 25/30 cycles
input lines IEC 61000-4-11
Single phase: at 0° 0%
UT; 250/300 cycles
Power frequency (50/60 Hz)
magnetic field
30 A / m
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Compliance level
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
Not applicable
Not applicable
Not applicable
30 A / m

Publicidad

loading

Este manual también es adecuado para: