Declaration – Electromagnetic Immunity
Immunity test
IEC 60601 test level
3 V
0.15 MHz to 80 MHz
Conducted RF
6 V in ISM bands
IEC 61000-4-6
between 0.15 MHz and
80 MHz
Radiated RF
10V / m
IEC 61000-4-3
80 MHz to 2.7 GHz
Compliance level
Immunity
test
Not applicable
10V / m
Radiated
RF
IEC
61000-
4-3
Note* – As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not
Declaration – IMMUNITY to proximity fields from RF wireless communications equipment
IEC60601 test level
Test frequency
Modulation
385 MHz
**Pulse Modulation: 18Hz
*FM+ 5Hz deviation:
450 MHz
1kHz sine
710 MHz, 745 MHz,
**Pulse Modulation: 217Hz
780 MHz
810 MHz, 870 MHz,
**Pulse Modulation: 18Hz
930 MHz
1720 MHz, 1845 MHz,
**Pulse Modulation: 217Hz
1970 MHz
2450 MHz
**Pulse Modulation: 217Hz
5240 MHz, 5500 MHz,
**Pulse Modulation: 217Hz
5785 MHz
represent actual modulation, it would be worst case.
Note** – The carrier shall be modulated using a 50 % duty cycle square wave signal.
Compliance level
Maximum
Immunity
power
level
1.8 W
27 V/m
27 V/m
2 W
28 V/m
28 V/m
0.2 W
9 V/m
9 V/m
2 W
28 V/m
28 V/m
2 W
28 V/m
28 V/m
2 W
28 V/m
28 V/m
0.2 W
9 V/m
9 V/m