Technical specifications

Accessories storage conditions
Accessories must remain in the packaging and be stored
within a temperature range of - 4° Fahrenheit to + 122° Fahrenheit
( - 20° Celsius to + 50° Celsius ).
Operation environmental conditions
The Invia Motion NPWT pump and accessories must be
operated within a temperature range of 41° Fahrenheit to
+ 104° Fahrenheit (+ 5° Celsius to + 40° Celsius ). Do not operate
the product in extreme cold or heat. Avoid high relative humidity
(15 – 93 % is permissable ).
CAUTION
Do not operate the Invia Motion NPWT pump above
9,842 feet (3000 meter) above sea level.
Protection Class IP22
The Invia Motion NPWT pump is protected against ingress
of solids larger than 0.50 inches (12.5 millimeter) and dripping water.