Electromagnetic compatibility (EMC)

Invia Motion is EMC-tested in conformity with the requirements of
IEC 60601-1-2:2007 and IEC 60601-1-2:2014 4th Edition according
clause 7 and 8.9.
WARNING
The Invia Motion pump should not be used adjacent to or
stacked with other equipment. If adjacent or stacked use is
necessary, the Invia Motion pump should be observed to
verify normal operation in the configuration in which it will be
used.
WARNING
Do not use other accessories than those specified or sold
by the manufacturer as replacement parts for internal com-
ponents as it may result in increased emissions or decrea-
sed immunity of the Invia Motion pump.
WARNING
Portable RF communications equipment (including periphe-
rals such as antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any part of the
Invia Motion. Otherwise, degradation of the performance of
this equipment could result.