The manufacturer cannot be held liable for accidental or indirect damages should the device be modified,
repaired without authorization or should any of its component be damaged due to accident or misuse.
Any minimal modification / repair on the device voids the warranty and does not guarantee the compliance with
the technical requirements provided by the MDD 93/42/EEC (and subsequent changes) and its normatives
- Before using the HOSPI PLUS SUCTION ASPIRATOR, consult the instructions for use: failure to read all the instructions in
this manual can be harmful for the patient.
- The device cannot be used to drain chest fluids;
- HOSPI PLUS SUCTION ASPIRATOR is not suitable for MRI. Do not introduce the device in MRI environments.
- The device must not be used for suction of explosive, corrosive or easily flammable liquids.
TYPOLOGY (MDD 93/42/EEC)
MODEL
UNI EN ISO 10079-1
POWER FEEDING
POWER CONSUMPTION
FUSE
MAXIMUM SUCTION PRESSURE (without jar)
MAXIMUM SUCTION FLOW (without jar)
WEIGHT
SIZE
DUTY CYCLE
SICILICONE TUBE SIZE
ACCURANCY OF VACUUM INDICATOR
WORKING CONDITION
CONSERVATION CONDITION AND TRASPORT
This product is guaranteed for a period of 24 months from the date of purchase. The warranty includes the repair or replacement of
defect spare parts free of charge, if the defect has been clearly described by the customer and determined by technical service.
Inspections on the part of the seller, performed at the request of the customer and intended to determined wjether the device is fully
functional, are not covered by the free-of-charge warranty service. This service will be charged to the customer depending on the
effort required. The consumables components are not subject to warranty. Consumable components are silicon tubes, filters, seals,
conical adaptor and suction catheter. Also exluded from warranty is all damage resulting from improper handling, wilful damage or
improper care of the device. The warranty shall expire if repairs and servicing are not carried out by technical service.
UNDER NEW EUROPEAN RULES, GIMA REQUIRES THE FOLLOWING PROCEDURES TO BE CARRIED OUT TO PROTECT THE
INSTRUMENT AND THE SAFETY OF ALL WHO COME IN CONTACT WITH IT.
Before returning an instrument for repair, the external surfaces and all accessories MUST be carefully disinfected with a cloth
soaked in methylated spirits or hypochlorite-based solution. The instrument and accessories should then be placed in a bag with a
note outlining the disinfection undertaken.
Instruments returned for repair MUST be accompanied by a description of the problem. GIMA will not be responsible for damage
Failure to follow this procedure will result in the instrument being returned to the purchaser unrepaired.
Where GIMA determines that an instrument is faulty, a replacement will be provided only if a SALES RECEIPT and STAMPED
caused through improper use. To avoid such damage, please read the instruction carefully.
GUARANTEE are provided. GIMA will not be responsible for damage accessories. These may be replaced at the customer's expence.
CONTRAINDICATIONS

TECHNICAL CHARACTERISTICS

Room temperature:
Room humidity percentage:
Atmospheric pressure:
Altitude:
Room temperature:
Room humidity percentage:
Atmospheric pressure:

WARRANTY CONDITIONS

RULES FOR RETURNING AND REPAIRING
14
Class IIa Medical Decice
HOSPI PLUS SUCTION ASPIRATOR
HIGH VACUUM / HIGH FLOW
230V~ / 50Hz
385 VA
F 1 x 4A L 250V
-90kPa / -0.90 Bar / -675mmHg
90 l/min
20 Kg
460 x 850 (h) x 420 mm
Non – Stop Operated
Ø 8 x 14 mm
± 5%
5 ÷ 35°C
30 ÷ 75% RH
800 ÷ 1060 hPa
0 ÷ 2000m s.l.m.
-40÷ 70°C
10 ÷ 100% RH
500 ÷ 1060 hPa