Using the footswitch control:
Connect the footswitch control feeding cable with the plug marked as
After the device has been connected, all Leds are still off. When the ON/OFF button is pressed, all Leds are activated at once for 1
second (autotest). At the end of the autotest cycle, the ON/OFF button led will flash. Press the button marked as (---) to perform
suction using the footswitch control and execute intermittence work cycles. Press the button marked as (-) to perform remote
suction using the footswitch control (continuous suction). To stop suction just push on the footswitch control with strength.
Using the footswitch control and the flow deviator: If using equipment fitted with a flow deviator, users may direct suctioned
liquids in any of the two collection tanks provided. Flow deviator comes with two complete suction kits (2 sets of tubes, 2
antibacterial and hydrophobic filters and two conical junctions).
After the device has been connected, all Leds are still off. When the ON/OFF button is pressed, all Leds are activated at once for 1
second (autotest). At the end of the autotest cycle, the ON/OFF button led will flash. To decide which side to perform the suction
from, press OUT LEFT or OUT RIGHT and the selected button led will show a blue light.
Press the ON/OFF button again to start the suction cycle. If the device is set up for using the flow deviator, ensure the antibacterial
filter has been positioned on both sides.
Connect the footswitch control feeding cable with the plug marked as
Press the button marked as (---) to perform suction using the footswitch control and execute intermittence work cycles.
Press the button marked as (-) to perform remote suction using the footswitch control (continuous suction). To stop suction just
push on the footswitch control with strength. Press the ON/OFF button to stop the medical equipment.
Before removing the feeding plug, ensure autotest has been performed on the panel.
Using FLOVAC® disposable collection system: Before connecting the disposable collection system, remove the blu ring fitted on
the tank holder for a more comfortable insertion of the same container.
After opening the package, fully stretch the bag and then flatten it concentrically to eliminate as much air as possible.
-
Insert the bag and apply the cover to an appropriately sized reusable rigid container by pressing firmly around the entire
-
perimeter. Make sure that the system is completely sealed.
Close the connector marked as "TANDEM" with the lid provided.
-
Connect the power source of the vacuum to the VACUUM port equipped with specific reusable conical fitting with "male"
-
connection.
Connect the patient tube to the PATIENT port of the cover
-
Before use, check all closures and make sure there are no leaks, starting the aspiration source. If the bag expands to fully adhere
-
to the walls of the rigid container and the cover bends towards the inside of the glass, the system is not leaking.
Start the aspiration and periodically check the filling level of the container. The overflow valve will cause the interruption of
-
aspiration if the aspirated fluids have reached the maximum filling level of the device.
When the float valve intervenes signalling the device is too full, the suction source must be disconnected within no more than 5
-
minutes. Warning: The accidental inversion of connections may cause contamination for the operator and/or for the
vacuum generation equipment.