FOOTSWITCH CONTROL (for versions equipped with footswitch)
Replacing the antibacterial filter:
The filter is made of hydrophobic material that stops the passage of liquids into the same filter.
If you suspect the filter may have been contaminated and/or got wet or discoloured, always remove and replace the filter.
If the equipment is to be used on patients with unknown pathological conditions or should you evaluate the possibility of indirect
contamination, remove and replace the filter after each utilization. The filter is not designed for decontamination, disassembly
and/or sterilization. If you suspect the filter may have been contaminated and/or got wet or discoloured, always remove and replace
the filter. If the equipment is to be used on patients whose pathologies are known and not implying any indirect contamination risks,
we recommend to remove and replace the filter at the end of each work shift or else every month, even if the equipment has not been
used. 4000ml (REF RE 210006) or 5000ml (REF RE 210010) complete jar versions are available on request.
Versions fitted with FLOVAC® 2000ml or 3000ml disposable collection systems (including a re-usable rigid polycarbonate container
and a disposable Liner) are also available on request.
Aspiration jar: The mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and sterilization. Beyond this
limit, the physical-chemical characteristics of the plastic material may show signs of decay. Therefore, we recommend that you to
change it.
Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the said tube.
Therefore, after each cleaning cycle, it is up to the final user to verify whether the tube is suitable for reuse. The component must be
replaced if there are visible signs of decay of the material constituting the said component.
Conical fitting: the number of cycles of sterilization and the number of cleaning cycles is strictly linked to the employment of the said
component. Therefore, after each cleaning cycle, it is up to the final user to verify whether the fitting is suitable for reuse.
The component must be replaced if there are visible signs of decay in the material constituting the said component.
Service life of the device: More than 10000-12000 hours of operation (or 3 years) in accordance with the standard conditions of
testing and operation. Shelf life: maximum 5 years from the date of manufacture.
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO
10993-1 standards on material biocompatibility. The medical device is provided without a specific suction probe. If this device
must be used with a specific suction probe, the end user is responsible for making sure it complies with the EN 10079-1 regulation.
Before using the device, the manufacturer advises you to clean and/or sterilize the accessories.
Washing and / or cleaning the autoclavable jar as to be carried out as follows:
Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating substances;
Disconnect the tank from the device and remove the said container from the support of the device.
Separate all the parts of the cover (overflow device, washer).
Disconnect all tubes from the jar and the protection filter
Wash each part of the container from secretions under cold running water and then clean every single part in hot water
(temperature not exceeding 60°C)
Once again, carefully wash each single part using, if necessary, a non-abrasive brush to remove any deposits. Rinse with hot
running water and dry all parts with a soft cloth (non-abrasive). It is possible to wash with commercial disinfectants by
carefully following the instructions and dilution values supplied by the manufacturer. After cleaning, leave the parts to dry
in an open, clean environment.
Dispose of the aspiration catheter according to that provided by local laws and regulations.
The silicone aspiration tubes and the conical fitting may be carefully washed in hot water (temperature must not exceed 60°C).
After cleaning, leave the parts to dry in an open, clean environment.
When cleaning is complete, reassemble the container for liquid aspirations according to the following procedure:
Place the overflow valve into its seat in the cover (under VACUUM connector)
Insert floating valve keeping the o-ring towards the opening of the cage
Place the o-ring into its seat around the cover
After completing assembling operations always make sue that cover seals perfectly to avoid vacuum leackages or liquid exit

ACCESSORIES SUPPLIED

ACCESSORIES SUPPLIED
N°2 COMPLETE ASPIRATION JAR 2000ml
CONICAL FITTING
SILICON SET TUBES 8 mm x 14 mm
ANTIBACTERIAL AND HYDROFOBIC FILTER
EUROPEAN POWER SUPPLY CORD
(H05VV-F - 2x0.75mm² - 2mt)
CLEANING ACCESSORIES AND INTERNAL PARTS
18
CODE
RE 210351/01
RE 210420
51100/01
SP 0121
Cod. 52130
SP 0020