36
Elektromagnetische
Verträglichkeit EMV
Guidelines and manufacturer's declaration – electromagnetic immunity
The ARTIP™ BASE holding system 28272 TLS and 28272 TSS is intended for use in an electromagnetic environment specified below. The user of the CMAC
Monitor 8403 ZX should make sure that it is used in such an environment.
Immunity tests
Electrostatic discharge (ESD) acc.
± 8 kV contact discharge
to IEC 6100042
± 15 kV air discharge
± 2 kV for power lines lines
Rapid transient electrical
± 1 kV for input and output
Interferences/bursts acc.
lines
to IEC 6100044
100 kH repetition
Surges
± 1 kV line(s) to line(s)
acc. to IEC 6100045
± 2 kV line(s) to earth
Voltage Dip:
Dip to 0% for 1 cycles @
0°phase angle
Dip to 70% for 25/30
cycles @ 0° phase angle
Voltage dips, short interruptions
and fluctuations of the supply
Dropout to 0% for 0.5
voltage acc. to IEC 61000411
cycles @ 0°, 45°, 90°,
135°, 180°, 225°, 270° &
315° phase angles
Voltage interruption:
100% for 250/300 cycles
Magnetic field at the
supply frequency (50/60 Hz)
30 A/m at 50Hz/60Hz
acc. to IEC 6100048
3V/m 80 MHz to 2.7 GHz
Radiated, radiofrequency,
electromagnetic field immunity test
*Refer table 2 for wireless Proximity
acc. to IEC 6100043
RF field test levels
Immunity to conducted
3V
disturbances, induced by radio
1kHz 80% AM modulation
frequency fields
6V
acc. to IEC 6100046
Electromagnetic
Compatibility EMC
Table 1
EN/IEC 60601 test level
Compliance level
± 8 kV contact discharge
± 15 kV air discharge
n/a
n/a
n/a
30 A/m at 50Hz/60Hz
10 V/m 80 MHz to 2.7 GHz
on 150 kHz to 80MHz
rms
n/a
in ISM bands
rms
Compatibilidad
electromagnética CEM
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
If image distortion occurs, it may be necessary to
position the ARTIP™ BASE holding system 28272 TLS
and 28272 TSS further from sources of power
frequency magnetic fields or to install magnetic
shielding. The power frequency magnetic field should
be me asured in the intended installation location to
assure that it is sufficiently low.
®