Switching on the lamp

1 Push the white switch to switch the white light on
(where applicable).
The green indicator light in the switch will come on.
2 Push the blue switch to switch the blue phototherapy
lamps on. The green indicator light in the switch will
come on.
It takes about 1 minute to reach maximum output.
3 The hour meter records the operating time of blue
phototherapy lamps.
If therapy is interrupted and the lights are turned of
Phototherapy lamps must be allowed to cool down for
about 1 minute before turning the back on.

Care of patient

Monitor child's core temperature.
Increase fluid supply, e.g to meet child's increased
fluid consumption, increase fluid supply by parenteral
infusion.
Check that eye protection remains in place at all
times.
Check serum-bilirubin concentration periodically.

Monitoring service life

The intensity of the effective light decreases with use.
After 500 operating hours the intensity is more than 85 %
and after 1000 operation hours the intensity is more than
80 % of the initial value.
Hours meter shows total operating time for the photo-
therapy lamp.
Blue phototherapy lamps must be replaced every
1000 operating hours by trained service personnel.
Use only the phototherapy lamps indicated on the Order
List. The use of other radiation sources can impair the
reliability and effectiveness of phototherapy.