Potentially negative effects on patients with
different underlying diseases
Medical device modification or misuse can be
dangerous.
Intended Use
Device for humidifying oxygen during inhalation or
insufflation therapy.
Used in combination with an O
flowmeter or O
2
DigiFlow for operation with O
cylinders or a central
2
supply system for Oxygen.
Only for patients with spontaneous breathing.
Safety information
WARNING
Risk of fire
Do not allow O
connections to come into
2
contact with oil, grease, or flammable liquids.
No smoking, no open flames.
WARNING
Risk of breakage
Make sure to install the breathing circuit
without loops and kinks.
CAUTION
Risk of personal injury and/or device malfunctions
Magnetic fields can impair the correct functioning of
the medical device and therefore endanger the
patient or user.
Do not use the medical device near MRI scanners.
12
Overview
B
-
2
A Housing
B Relief valve
C Sealing ring
D Hose 5 x 2 Silicone 60 Sh A, 2 cm
E Set bubbler
F Humidifier bottle
Preparation
Before using for the first time
A
CAUTION
Risk of breakage
Check the medical device before use. Do not use
the medical device if it is damaged.
Make sure that the individual parts have been
C
disinfected/sterilized, see page 14.
D
E
Assembling humidifier
F
Instructions for use Humidifier Bubbler
A
B