07 . dELEtE FuNCtIONS
the following functions can be found in the Menu system:
Delete current data:
If you determine that the last measured value was
incorrect, it can be deleted using this function . The
statistics are updated at the same time .
Delete all data:
All data in the respective working mode can be dele-
ted using this function .
Delete group data:
This function includes the function "Delete all data" .
In addition, the upper limit value, the lower limit value
and the single and two-point calibration values are
deleted .
08 . CONtrOLLINg thE MEASurINg
INStruMENt WIth thE PC
The measured values of all working modes can be
downloaded to the PC via the USB interface . Further
information can be found in our software manual .
The free software supplied is designed to pro-
vide useful basic functions. Trotec provides
no guarantee whatsoever on this free soft-
ware and also does not provide support for it.
Trotec accepts no liability for the use of the
free software and is under no obligation to
make corrections or to develop updates and
upgrades.
B - 10
Operating manual – layer thickness measuring instrument BB 20
09 . trOuBLEShOOtINg
The following list of error messages shows you how
to recognise and rectify faults .
Error code
Error
Err1, Err2, Err3
The sensor is incorrectly con-
nected;
Losses; deviating signal
Err1
Eddy current sensor
Err2
Magnetic induction sensor
Err3
Both sensors
Err4, Err5, Err6
Reserved
Err7
Error in the layer thickness .
G