Maintenance
9. Open the test gas cylinder valve to let gas flow over the sensor
with a volume flow of 0.5 l/min.
The displayed, flashing measurement value changes to the value
according to the supplied test gas.
When the displayed measurement value is stable (after at least
120 seconds):
a. Press the OK-key to carry out the calibration.
The display containing the current gas concentration changes
with the display OK.
b. Press the OK-key or wait for approx. 5 seconds to end the
calibration/adjustment of this measuring channel.
Potentially, the next measuring channel will be offered for
calibration.
After the calibration/adjustment of the last measuring channel,
the instrument changes to the measuring mode.
c. Close the test gas cylinder valve and remove the instrument
from the calibration cradle.
If a fault occurred during the span calibration/adjustment:
The fault message
appears and
respective sensor instead of the measured value.
In this case, repeat the calibration/adjustment.
If necessary, change the sensor.
Notice for the adjustment of the ex-channel to nonane as
a measuring gas:
During the calibration of the ex-channel, propane can be used as
a substitute calibration gas.
When using propane to adjust the ex-channel to nonane,
the display must be set to twice the used test gas concentration.
Notice for the use in subsurface mining:
For the calibration of the ex-channel to the measuring gas
methane, the display of the instrument must be set to a value of
5 % (relative) lower than the used test gas concentration.
26
5
5.1
The instrument should be inspected and maintained annually by suitably
qualified persons (see: EN 60079-29-2 – Guidelines for selection,
installation, use and maintenance of instruments for the detection and
measurement of flammable gases and oxygen. EN 45544-4 – Electrical
instruments for the direct detection and direct concentration
measurement of toxic gases and vapours – Part 4: Guide for selection,
installation, use and maintenance and national regulations).
Recommended calibration interval for measuring channels Ex, O
H
S and CO: 6 months. Calibration intervals of other gases: see
2
Instructions for Use of the respective DrägerSensors.
5.2
The instrument does not need any special care.
Dirt and deposits can be removed from the instrument by washing
is displayed for the
it with cold water. A sponge can be used for wiping if necessary.
i
i
Carefully dry the instrument with a cloth.
Maintenance

Maintenance table

Cleaning
NOTICE
Abrasive cleaning tools (brushes etc.), cleaning agents and
cleaning solvents can destroy the dust and water filters.
,
2