1 Output format in TOLEDO Continuous Mode

Weight values are always transmitted in the following format:
STX
SB1
STX
ASCII characters 02 hex/2 deci, character for "start of text"
SB...
For status bytes, see below
DF1
Data field with 6 digits for the weight value transmitted without a decimal point and
unit
DF2
Data field with 6 digits for the tare weight
CR
Carriage return (ASCII character 0D hex/13 deci)
CHK
Checksum (2-part complement of binary sum of 7 lower bits of all previously
transmitted characters, including STX and CR)
Status byte SB1
Bit 6
Bit 5
0
1
Bit 4
Bit 3
0
1
1
0
1
1
SB2
SB3
Bit 4
Rounding / Increment
Rounding/
Increment
1
2
5
DF1
Bit 3
Bit 2
Bit 2
Bit 1
0
0
0
0
0
1
0
1
1
0
1
0
1
1
1
1
157
DF2
CR
Bit 1
Decimal position
Bit 0
Decimal
position
0
XXXX00
1
XXXXX0
0
XXXXXX
1
XXXXX.X
0
XXXX.XX
1
XXX.XXX
0
XX.XXXX
1
X.XXXXX
CHK