Work Sequence: New Measurement - Hach LINK2SC Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 28
2. Process the job with the photometer (refer to
SD memory card is being used, it must be connected to the photometer via a USB adapter.)
a. Use the LINK2SC button on the toolbar to open the job list. Pending jobs are identified with
numbers in a yellow circle.
b. Select a job.
c. Select the parameter to be processed.
d. Analyze the comparison sample.
e. Assign the result to the job, if applicable.
f. Use SEND to adjust. (If the SD memory card is being used, it must be connected to the
photometer via a USB adapter.)
4
3. LAN
: When the default setting is active, the probe is automatically corrected with the laboratory
measurement data once the job has been returned.
Note: LINK2SC>CONFIGURE>AUTO JOB can be used to deactivate automatic correction. The probe must
then be corrected manually as per transfer via SD memory card.
or
SD memory card: The job must then be activated manually on the controller. Proceed as follows:
a. Insert the SD memory card in the sc controller.
b. Select LINK2SC from the sc controller menu.
c. Select JOBS FROM CARD from the LINK2SC menu. (Jobs are loaded into the sc controller.)
d. Select JOB LIST from the LINK2SC menu.
e. Select the corresponding job from the list.
f. Select ACTIVATE JOB. (The probe is corrected.)

Work sequence: NEW MEASUREMENT

1. In order for LINK2SC to be able to select the best correction type for the application in question
(refer to
Introduction to matrix correction
performed via "NEW MEASUREMENT" and not via "CREATE JOB".
a. Select LINK2SC from the sc controller menu.
b. Select JOB LIST from the LINK2SC menu.
c. Select the corresponding job from the list.
d. Select NEW MEASUREMENT.
e. Select the PARAMETER for the new measurement.
f. Take a comparison sample for the laboratory measurement.
g. Use JOB->LAB to transfer the job file from the job status menu to the laboratory. (SD memory
card must be inserted in the sc controller if this is being used.)
2. Process the job with the photometer (refer to
5
3. LAN
: The probe is automatically corrected with the laboratory measurement data once the job
has been returned.
or
SD memory card: The job must then be activated manually on the controller (refer to
sequence: CREATE JOB
4
Only valid for sc1000 controllers.
5
Only valid for sc1000 controllers.
6 English
Work sequence: photometer
on page 4), each further sensor correction must be
Work sequence: CREATE JOB
on page 5 step 3).
on page 8). (If the
on page 5 step 2).