Operating Instructions; Connect Basic To Mains Power; Check Maximal Vacuum For Functional Check - Medela Basic Instrucciones De Uso

Ocultar thumbs Ver también para Basic:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 258

7 Operating instructions

CAUTIONS
The Basic is to be set up in such a way, that a separation from the mains supply can
be easily managed.
7.1
Connect Basic to
mains power
7.2
Check maximal
vacuum for functional
check
Specifications:
Altitude
above
sea level:
Max. Vacuum:
+ 2000 m
– 525 mmHg
+ 1000 m
– 592 mmHg
+ 500 m
– 638 mmHg
0 m
– 675 mmHg
(Tolerance: +/- 15 %)
18 |
Operating instructions
1.
Check the pump before use following the instruction in
chapter 6.1.
2.
3.
1.
2.
3.
– 70 kPa
– 79 kPa
– 85 kPa
– 90 kPa
2.1 Connect the power cord
to the appliance inlet at
the back of the Basic.
2.2 Plug in the mains plug of
the power cord to a fixed
mains socket.
3.1 An internal self-test is
performed. When the
green LED lights up, the
device is ready for use.
1.1 Switch on the Basic.
2.1 Turn the vacuum regulator
to the right to set
maximum vacuum
3.1 Seal the end of the patient
tubing with your thumb.
3.2 Compare the maximum
vacuum according to the
specification (on left). See
chapter 11 if the maximum
vacuum is not reached.

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido