6 Preparation for use

WARNINGS
For use only by medically trained persons who have been adequately trained in suction
procedures and in the use of aspirators.
CAUTIONS
– The Basic must remain in an upright position during use.
– The rack version requires a minimum distance of 5 cm to the enclosure to prevent from
overheating of the device. The back of the enclosure must be open.
– Sterile accessories must be checked to ensure the integrity of the packaging before
use.
– Non sterile and reusable accessories must be cleaned, disinfected and / or sterilized
according to the Medela cleaning guide (product code 200.2391).
6.1

Checks before use

– Check the Basic system before use for damage of the power cord or plug,
obvious device damage or safety defects and proper functioning of the device.
– Check for completeness and general condition of the Basic delivery package.
– Check all accessories prior to use:
– suction jars, lids and liners for cracks, brittle and flawed spots. Replace if
necessary.
– tubing for cracks, brittle areas and that connectors are firmly attached.
Replace if necessary.
– as an additional safety test, evacuate the system (including jars) to maximum
vacuum before actual use.
6.2
Assembly of the
basic configuration
1.
2.
3. Attach all necessary accessories according to your needs.
See annex "II List of accessories".
1.1 Make sure that the Safety
Set is attached to the
Basic. See chapter 5.5.
2.1 Attach a filter (if required)
to the Safety Set with the
arrow pointing in the flow
direction.
Preparation for use