8 Preparation for use / operating
instructions
WARNINGS
For use only after instruction by medically trained and qualified persons who have been
adequately trained in the use of Thopaz. Wear gloves for all operations.
«Pressure» generally implies «negative pressure».
CAUTIONS
– Sterile accessories must be checked on the integrity of the packaging before use.
– Non sterile and reusable accessories must be clean and disinfected according to
the Medela cleaning guide 200.2391
– Do not switch Thopaz on before charging the battery first.
8.1

Checks before use

– Check Thopaz system before use for damage of the power cord or plug, device
damage or safety defects and proper functioning of the device.
– Check that the seal is correctly placed in the vacuum port of Thopaz. If you do not
know where the vacuum port is placed please refer to Chapter 5 – Overview.
– Check for completeness and general condition of the Thopaz delivery package.
– Make sure that the rechargeable battery is recharged in case Thopaz is operated in
battery mode.
– Check all accessories prior to use:
– canister for cracks, brittle and flawed spots.
– tubing for cracks, brittle areas and that connectors are firmly attached. Replace if
necessary.
8.2

Connect tubing

22 |

Preparation for use / operating instructions

1.
2.
1.1 Open external packaging.
2.1 Insert the smaller of the
two connectors (connec-
tion to pump) horizontally
in the direction of the
arrow.