Record The Installation Information; Install A Desiccant Cartridge - Hach Flo-Station Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Figure 9 Relay, output power and datalog connections
1 Common contact for the relay (CCW)
2 Normally-open contact for the relay (CCA)
3 +12 VDC for the output power connection
4 –12 VDC (ground) for the output power connection

Record the installation information

Record the installation information so it can be entered in the Flo-Ware software:
• Site specific data (location, site ID and description)
• Installation date
• Pipe inside dimension (ID) number
• Level offset, Sensor Offset or Level Cal as appropriate. Refer to the sensor manual for the
installed sensor.
• Sediment level if applicable
Startup

Install a desiccant cartridge

If the instrument is installed in a high-moisture environment, install a desiccant cartridge to absorb
moisture. Refer to
Replacement parts and accessories
The desiccant cartridge prevents damage to the APR tube from moisture and debris. Moisture and
debris decreases the accuracy of the surcharge level pressure transducer on a Flo-Dar sensor or
primary level on a Flo-Tote 3 sensor.
If the instrument is installed in a relatively dry environment, a desiccant cartidge is not necessary.
The hydrophobic filter installed in the APR port is all that is necessary to absorb moisture.
1. Pull the hydrophobic filter off the APR port. Refer to
2. Push the desiccant cartridge tube on the APR port.
5 Input for the data logging connection (VCH+)
6 Common for the data logging connection (VCH-)
7 Not used
8 Not used
on page 19.
Figure
10.
English 15

Publicidad

Capítulos

Tabla de contenido

Solución de problemas

loading

Tabla de contenido