The maximum approved body weight is specified in the technical data
(see Page 20).
2.3 Environmental conditions
Storage and transport
Temperature range –20 °C to +60 °C (–4 °F to +140 °F), relative humidity 20 % to
90 %, no mechanical vibrations or impacts
Allowable environmental conditions
Temperature range: –10 °C to +45 °C (14 °F to 113 °F)
Chemicals/liquids: fresh water, salt water, perspiration, urine, soapsuds, chlorine
water
Moisture: Submersion: max. 1 h in 3 m depth, relative humidity: no restrictions
Solids: Dust, occasional contact with sand
Clean the product after contact with humidity/chemicals/solids, in order to
avoid increased wear and damage (see Page 19).
Unallowable environmental conditions
Solids: Highly hygroscopic particles (e.g. talcum), continuous contact with sand
Chemicals/liquids: Acids, continuous use in liquid media
2.4 Service life
Prosthetic foot
This product has been tested by the manufacturer with 2 million load cycles
according to ISO 22675. Depending on the patient's activity level, this cor
responds to a service life of 2 to 3 years.
Footshell, protective sock
The product is a wear part, which means it is subject to normal wear and
tear.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or injury.
CAUTION
Warning regarding possible technical damage.
NOTICE
3.2 General safety instructions
CAUTION
Exceeding the service life and reuse on another patient
Risk of injury due to loss of functionality as well as damage to the product
14