Servicing & Repair; Fuse Replacement; Limited Warranty; End Of Product Life - Sunoptic Surgical LED-4000 Manual Del Operador

Fuente de luz led
Ocultar thumbs Ver también para LED-4000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 58
8. SERVICING AND REPAIR
Defective items are to be serviced and repaired exclusively by persons authorized by the manufacturer.
All repair work shall employ original manufacturer's parts only.

8.1 FUSE REPLACEMENT

Always disconnect power cord and turn main switch off before fuse
replacement.
CAUTION
Turn light source off and unplug power cord. On back of unit remove the fuse holder located in the
power cord receptacle and pull out fuse holder. Users should use 2A Fast-acting 250VAC rated fuses.
Insert holder back into fuse housing.
Re-connect the power cord and turn the LED light source on according to section 6.1.

8.2 LIMITED WARRANTY

Your LED-4000 light source carries a 3-year warranty from the date of shipment on workmanship and
all defects of material. Should your product prove to have such defects within three years from the date
of shipment from Sunoptic Technologies
without charge. Should your LED-4000 light source need servicing under this warranty, please contact
Sunoptic Technologies
a sturdy carton and ship it to the factory. Please include a note describing the defects, your name,
telephone number and a return address. Warranty does not cover equipment subject to misuse,
accidental damage, normal wear and tear or if transferred to a new owner without authorization from
Sunoptic Technologies
rights that vary from state to state.
POST WARRANTY REPAIRS: You may return your product(s) for repair, shipping prepaid to the factory.
Your product will be inspected and an estimate of repair charges will be submitted to you for approval.
Payment must be received before repairs are completed.
In the US:
FAX number:
Customer Service
9.

END OF PRODUCT LIFE

In accordance with the European Waste from Electrical and Electronic Equipment (WEEE) directive,
we encourage our customers to recycle this product whenever possible. Disposal of this unit must be
performed in accordance with the applicable local environmental regulations.
In the US a list of recyclers in your area can be found at: http:/www.eiae.org/.
Please contact our Repair Department to issue a return authorization to return product to manufacturer
at the end of product life.
LIT-216 Sunoptic Surgical
Rev. C
®
, we will repair or replace the product or component part
®
for return authorization documentation. You should carefully pack the unit in
®
.This warranty gives you specific legal rights and you may also have other
®
Page 12 of 114
877-677-2832 (toll free)
904 733-4832
904-737-7611
(English)

Publicidad

Tabla de contenido
loading

Tabla de contenido