Maintenance, Servicing, Repair & Warranty; Fuse Replacement; Warranty; Repair - Sunoptic Surgical SSL-2000 Manual Del Operador

Ocultar thumbs Ver también para SSL-2000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 49
WARNING: The illuminator is not sterilizable. DO NOT attempt to autoclave the device.
8. MAINTENANCE, SERVICING, REPAIR & WARRANTY

8.1. Fuse Replacement

WARNING: Always TURN OFF and UNPLUG the illuminator from mains power before attempting to
replace a fuse.
Turn off illuminator and unplug power cord. Remove the defective fuse by rotating the fuse holder
cover and extracting the fuse from the holder. Replace blown fuse with ¼" x 1-1/4" (6.4 x 31.7mm),
250V, 2A, Time Delay fuse. Reinstall fuse holder cover.

8.2. Warranty

The illuminator carries a 3-year warranty from the date of shipment on workmanship and all defects
of material.
Should your product prove to have such defects within three years of shipment, Sunoptic
®
Technologies
will repair or replace the product or component part without charge. Should your
product(s) need servicing under this warranty, please contact Sunoptic Technologies
distributor for return authorization documentation.
Please carefully pack the unit in a sturdy carton and ship it to the factory. Please include a note
describing the defects, your name, telephone number and a return address. Warranty does not
cover equipment subject to misuse, accidental damage, normal wear, and tear or if transferred to
a new owner without authorization from Sunoptic Technologies
legal rights and you may also have other rights that vary from state to state.

8.3. Repair

You may return your product(s) for repair, shipping prepaid to the factory. Your product will be
inspected, and an estimate of repair charges will be submitted to you for approval.
PHONE: +1 (877) 677-2832
FAX: +1 (904) 733-4832

8.4 Troubleshooting

Problem
The power indicator (refer to 5.) is
not lit.
The power indicator is lit but the
lamp will not ignite.
9.

END OF PRODUCT LIFE

In accordance with the European Waste from Electrical and Electronic Equipment (WEEE) directive,
we encourage our customers to recycle this product whenever possible. Disposal of this unit must be
performed in accordance with the applicable local environmental regulations.
In the US a list of recyclers in your area can be found at: http:/www.eiae.org/.
Please contact customer service to issue a return authorization to return product to manufacturer at
the end of product life.
LIT-168
Sunoptic Surgical
Rev. H
INTERNATIONAL: +1 (904) 737-7611
Check that the AC power cord is properly connected.
Check the unit fuses and replace if necessary.
Turn intensity control knob clockwise to increase light output
intensity.
®
Page 11 of 96
®
. This warranty gives you specific
Solutions
®
or a local
(English)

Publicidad

Tabla de contenido
loading

Tabla de contenido