Menu (details)
Stirring
Speed Limit:
The "Speed Limit" menu allows the user to set the desired maxi-
mum upper speed limit for the EUROSTAR 200 / 200 P4 control
stirrer. The
initial setting is the maximum permissible speed of
the stirrer. If the user changes this setting, the Wireless Controller
saves the new value for future stirring tasks.
If the "Speed Limit" has been changed, then the speed can be
adjusted only within the new range.
Torque Limit:
The "Torque Limit" menu allows the user to set the desired maxi-
mum deliverable torque limit. The initial setting is the maximum
permissible torque of the device.
If the user changes this setting, the Wireless Controller saves the
new value for future stirring tasks.
If the "Torque Limit" has been changed, the stirrer can deliver an
operating torque only up to the value specified as the maximum
torque limit.
Note: The torque limit can be overload for about 10 seconds. This
is necessary to cater for stirring tasks which require metering and
feeding additives.
Intermittent Mode:
Symbol rotation
Chuck rotation
direction
direction
CW
CW
Fig. 9
Torque Display:
The "Torque Display" menu allows the user to specify that the
torque is shown in the display. A tick shows that the option is
activated.
Note: Pressing the "Back" button while the device is running
resets the torque to 0 Ncm. The Δ symbol then appears in the
display in front of the torque value.
Torque Calibration:
This menu allows the torque to be calibrated. All torque resulting
from bearing friction is excluded from the torque calculation. This
is performed with no stirrer element attached. A duration of 30
sec. and speed of 50 rpm are set automatically.
Note: Calibration can be performed only when a USB cable is be-
ing used. For this, remove the Wireless Controller from the sta-
tion (see "Interfaces and outputs" section, figure "Connection
capability: Wireless Controller to the EUROSTAR station").
Time
∞
/ CW...
Time
Run-Stop-Run / CW ...
Graph
Factory setting: continuous mode
Function "Run / Stop" is activated:
•
The run time and stop time can be set separately.
28