Electromagnetic Compatibility; Electromagnetic Emissions - EMS Piezon Master 400 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Page E-13/E-19

Electromagnetic compatibility

Electromagnetic emissions

The Piezon Master 400 is intended for use in the electromagnetic environment specified below. The
customer or the user of the Piezon Master 400 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage
fluctuations/Flicker
emissions
IEC 61000-3-3
Compliance
Electromagnetic environment - guidance
Groupe 1
The Piezon Master 400 uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
Class B
The Piezon Master 400 is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
Class A
purposes.
Complies