Technical data
Degree of protection
Electromagnetic compatibility
Emitted interference and noise immunity
Display and controls (see Figure 3, page 46)
Display
Resolution with maximum measurement range
Position of decimal point
Limits
Display of limit violation
Parameter adjustment
Units
Damping
Design
Weight
Materials
Materials of non-wetted parts
Field housing
Cover
Materials of wetted parts
Protective tube
Threaded plugs on the protective tube
Measuring insert
Connection from the display to the
protective tube
Length of the protective tube (U
Lagervarianten
44
IP65 to EN 60 529
In accordance with EN 61 326 /A2 Annex A (2001),
see declaration of conformity in the Certificates
chapter
LCD, max. 5 digits, digit height 9 mm
0.01 °C (0.01 °F)
Programmable
Programmable
Red LED and signal on LCD (↑ symbol / ↓ symbol
for upward or downward violation)
via 3 buttons
mA or % or physical variable:
°C, °F, °R, K
Between 0.1 and 100 s
(step size: 0.1 s) programmable
About 0.7 kg (0.70 kg)
∅ 80 mm (3.15 in), stainless steel
(material no. 1.4016)
Stainless steel (material no. 1.4016) with glass
in accordance with DIN 43 772, form 8
(March 2000),
∅ 14 x 1.5 mm (∅ 0.55 x 0.06 in)
Stainless steel (material no. 1.4571 / 316Ti)
G½B in accordance with DIN 3852-2 Form A,
stainless steel (material no. 1.4571/316Ti),
or 1/2" - 14 NPT stainless steel (material no.
1.4571/316Ti)
Length to match the protective tube ordered, stain-
less steel
- Radial (type A), can be pivoted by max. ±120° (α)
- Axial (type B), can be pivoted by max. 360°
)
73 mm ... 2500 mm (2.87 ... 98 inch)
1
170 mm (6.70 inch)
260 mm (10.24 inch)