Servicing and maintenance
After opening the device, there is a danger of influencing the signal by touching the
electrical connections. This can be avoided by switching off the supply voltage or sep-
arating the signal circuit.
After installation has been successfully completed, the measuring insert with Pt100
housing unit can be dismantled and reinstalled on site. In order to do this, loosen the
screw cap (1, Figure 4, p.50). For normal calibration runs, no separation of the con-
nection line is necessary.
Function
Mode
↑
Measured
(1)
Modify
value
measured
value
display
Password
2
higher
Calibrate out-
3a
higher
put current to
4 mA
Calibrating
3b
higher
the output cur-
rent to 20 mA
factory setting 3d
run "↑" and "↓" simulta-
neously: and ⇒ M1
Table 2
Service parameters
SITRANS TF2
A5E00263328-02
Button function
↓
M
⇒ M2
Modify deci-
mal position
"99" - ⇒
lower
M4
"105" - ⇒ M3a
"110" - ⇒ M3d
short ⇒
lower
M3b
long ⇒
M1
short ⇒ M3a
lower
long ⇒ M1
⇒ M1
Display,
explanatory comments:
Measured value in mA or % or in
physical unit °C,°F, °R and K
Enter password
"99" – Parameters (see page 40)
"105" – Calibrate the output current
"110" – Reset to default setting
Output current in mA,
if "↑" or "↓"
is pressed briefly, → Slight
change, otherwise rapid change
speed
Output current in mA,
if "↑" or "↓" are pressed briefly.
→ Slight change, otherwise rapid
change speed
Reset to factory setting
5