Varian Turbo-V 150 Manual De Instrucciones página 67

Tabla de contenido

Publicidad

Enter the value by pressing the PUMP CURRENT
pushbutton, and the display shows:
R S 2 3 2
H O S T
S E L E C T
I O N :
where: XXXX == means HOST or PRINT.
Select HOST or PRINT by pressing the CYCLE
pushbutton.
R S 2 3 2
H O S T
S E L E C T
I O N :
/
P R I
N T
X X X X X
/
P R I
N T
H O S T
62
TECHNICAL INFORMATION
With
the
RS
232
communication is established by selecting HOST. Data
are sent to an external computer every time the
external computer asks for the values.
The data available are:
Pump/controller operating condition
Cycle time
Pump life
Pump temperature
Pump current
Pump voltage
Controller output frequency
Cycle number
R1 condition
R2 condition
Life time
Cycle # zeroing
Configuration parameter readings
Configuration parameter setting
If PRINT is selected and a printer is connected on RS
232 line, a unidirectional communication is established
and every minute the data are sent to the printer, even
if the pump is not running.
The set of data available are:
Pump speed KRPM
Pump temperature
Pump current A
Pump power W
R1 condition
R2 condition
The controller is factory-set to HOST.
Confirm the selection by pressing the PUMP
CURRENT pushbutton.
In this way you enter into an operating phase
named "Monitor Relay Programming" described in
the following paragraph.
connected,
a
bidirectional
87-900-907-01 (B)

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

969-9436969-9536

Tabla de contenido