Operating the Device - General
The insufflation mode display depicts the selected mode (e. g. Mode: High Flow).
EN
Stop insufflation:
The STOP function field is displayed while insufflation is active.
Stop insufflation by pressing the STOP function field.
The status line depicts Insufflation stopped followed by the selected mode (e.g.
Insufflation: High Flow) alternating with Push
to release. The insufflation
mode display depicts Change Mode.
In case the tube set connection is lost, the insufflation stops and the display de-
picts Insufflation stopped ->Mode: High Flow <->Tube not connected.
The device depicts the same message if a tube set has not been connected or has
not been inserted correctly and the START function field has been pressed. In this
case the insufflation will not be started. Please insert a tube set or check the tube
set connection.
5.6.7

Using the SIDNE Port (Optional)

The SIDNE Port allows connection to the SIDNE control system. When connected,
all of the functions and settings of the insufflator can be:
• controlled with SIDNE voice and pendant commands; and
• displayed on the SIDNE video overlay and pendant.
The physician(s) and support personnel must be thoroughly familiar with the set-
up and operation of the SIDNE. Please refer to the SIDNE documentation for the
proper setup, use and troubleshooting. When SIDNE is connected and active, the
functions and settings of the insufflator may simultaneously be adjusted using
the buttons on the front panel and the SIDNE.
5.6.8

Turning Device Off

Use the ON/OFF switch to turn the device off. The device is turned off.
25