Selecting Vessel Harvest Operating Mode - Stryker PNEUMO SURE Instrucciones De Uso

Ocultar thumbs Ver también para PNEUMO SURE:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Using and Controlling the PNEUMO SURE XL High Flow Insufflator in Vessel Harvest Operating Mode
EN
(1)
(2)
52
expressed with the following symptoms:
• Reduced respiration with restricted diaphram function
• Hypercapnia
• Reduction of venous reflux
• Reduced cardiac output
• Metabolic acidosis
WARNING!
Dehydration
Insufflation can lead to dehydration of the tissue. This can result in organ tissue
damage and cardiovascular reactions of the patient. Long surgeries and large
leaks increase the risk of dehydration (especially at the insertion points of the
trocars or when changing instruments).
WARNING!
Embolism
Improper placement of the insufflation instrument could cause insufflation of
gas into a vessel, resulting in air or CO2 embolisms. To reduce the risk of air or
CO2 embolism, perform initial insufflation at a low flow rate and ensure that the
insufflation instrument is correctly positioned. Check the position of the insuf-
flation instrument immediately if the actual pressure rapidly reaches the nomi-
nal pressure value. CO2 embolisms can also be caused by a high pressure. Avoid
high-pressure settings and close damaged blood vessels at once.
WARNING!
Device-inherent dangers
Please read the general risks and dangers information in chapter 2.1 Hazards,
page 5 and chapter 3.3 General Device-Inherent Dangers, page 8.
9.2

Selecting Vessel Harvest Operating Mode

1. Selection of the insufflation mode can be done only while insufflation is
stopped. If Vessel Harvest operating mode is not yet set and displayed, press
the function field Change Mode for 2 seconds to get the insufflation mode
overview.
2. The display depicts an insufflation modes overview. Select the desired insuf-
flation operating mode by pressing the Vessel Harvest function field.
3. The procedure profile is depicted on the display. The displayed parameters
correspond with the factory settings or the values set in the Configuration
menu (see chapter 10.1 Configuration menu I, page 60).

Publicidad

Tabla de contenido
loading

Tabla de contenido