Presetting Nominal Flow In Pediatric Operating Mode - Stryker PNEUMO SURE Instrucciones De Uso

Ocultar thumbs Ver también para PNEUMO SURE:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

the intra-abdominal pressure ends. Pressing the nominal pressure ºfunction field
again does not increase the pressure any further. The status line of the display
depicts Safety limit and the nominal value flashes.
Release the function field at that point. The display switches back to the nominal
value setting after 2 seconds. Now you can set a value up to 20 mm Hg.
7.4

Presetting Nominal Flow in Pediatric Operating Mode

The nominal flow can be set during insufflation or while insufflation is stopped.
Reducing nominal gas flow:
Press the nominal gas flow function field » to reduce the nominal gas flow. The
nominal gas flow is decreased in the range of
• 0.1 l/min to 2 l/min in increments of 0.1 l/min.
• 2 l/min to 20 l/min in increments of 1 l/min.
Keeping the » function field depressed longer than 1.5 seconds activates scroll-
ing.
• In the range of 0.1...2 l/min, rounding down to 1 or 0.1 l/min is applied.
• In the range of 2...20 l/min in increments of 1.
The nominal flow can be set during insufflation or while insufflation is stopped.
Increasing nominal gas flow:
Press the nominal gas flow function field º to increase the nominal gas flow. The
nominal gas flow is increased in the range of
• 0.1 l/min to 2 l/min in increments of 0.1 l/min.
• 2 l/min to 20 l/min in increments of 1 l/min.
Keeping the º function field depressed longer than 1.5 seconds activates scroll-
ing.
• In the range of 0.1 l/min...<2 l/min, rounding up to 1 or 2 l/min is applied.
• Then in one-step increments down to 20 l/min.
Safety limit:
When increasing the nominal flow, the status line of the display depicts Safety
limit starting at 5 l/min and the nominal value flashes. (This safety limit can be
activated or deactivated in the Configuration menu).
Release the function field to increase the nominal gas flow further. The display
switches back to the nominal value setting after 2 seconds. Now you can set a
value of up to 20 l/min.
WARNING!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
Using and Controlling the PNEUMO SURE High Flow Insufflator in Pediatric Operating Mode
Set Pressure
Set Pressure
Actual Pressure
Actual Pressure
Set Flow
Set Flow
mm Hg
mm Hg
mm Hg
mm Hg
l/min
l/min
0
15
0.1
283
Liter
Safety limit
Safety limit
Change
Actual
START
Mode
0
l/min
EN
37