13.6
Max. Device Pressure Test
(1)
mm Hg
(4)
1. Set to High Flow operating mode (Pneumo Sure XL: Set to Bariatric operating
mode).
2. Select the max. nominal gas flow.
3. Connect a manometer (1) and an open Veress cannula (4) to the insufflation
tube connection (3)/adapter (2).
4. Select the max. gas flow.
5. Start insufflation:
Press the START function field. The manometer registers a pulsing pressure
increase. When the pressure stabilizes, the manometer registers a maximum
pressure of 55-65 mm Hg (Bariatric: 65-75 mm Hg).
6. Stop insufflation:
Press the STOP function field.
13.7
Gas Flow Rate Test
Test setup with open connection, without connected insufflation tube.
• Select a nominal gas flow rate of 15 l/min.
• Start insufflation:
Press the START function field.
• Press the gas volume function field (0.0 l must be displayed).
Now start measuring for one minute.
• Stop the insufflation after one minute: Press the STOP function field.
The gas consumption should be at least 11-12 l.
Each successfully conducted test must documented with the test log.