Overview Notifications - Medela Invia Motion Instrucciones De Uso

Ocultar thumbs Ver también para Invia Motion:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 99

Overview notifications

Visual
Notification
symbol
Description
on
display
Leak in
system
Blinking!
+
acoustic
signal!
Blockage
Blinking!
+
acoustic
signal!
Canister full
Blinking!
+
acoustic
signal!
Battery low
Blinking!
+
acoustic
signal!
Battery fully
discharged
Blinking!
+
acoustic
signal!
Pump time
expired
Blinking!
(Exam-
ple for
15 days
version)
24
Acoustic signal
Troubleshooting
Can be muted
Check the system
for 60 seconds
for leaks.
by pressing
Can be muted
Check the system
for 60 seconds
for blockage.
by pressing
Verify that the
correct canister is
used:
Can be muted
Change the Invia
for 60 seconds
Motion canister/
by pressing
tubing set.
Can be muted
Connect the Invia
permanently by
Motion pump to
pressing
the power supply
and fully charge
battery. Charging
time can be up to
4 hours.
Can be muted by
Connect the Invia
switching off the
Motion pump to
pump.
the power supply
and fully charge
battery. Charging
time can be up to
4 hours
The blinking
Dispose of the
signal can be
Invia Motion NPWT
suppressed
pump. Attach
permanently by
another device if
pressing
necessary.
Remarks/Potential
cause
Dressing could be
leaking, check that it is
properly sealed.
The dressing or canister
tubing could be twisted,
kinked or clamped. If
the Invia Motion pump
detects blockage in the
087.4016.
intermittent mode, it will
continue to run in the
on-state until blockage
is resolved.
The filter could get wet
by improper positioning
of the pump (upside
down or display side
down).
The visual symbol blinks
all the time until battery
is fully discharged. The
remaining time of the
battery is approximately
30 minutes.
The remaining time
of the battery is approxi-
mately 10 minutes.
The Invia Motion NPWT
pump provides an addi-
tional safety usage time
of 72 hours. During this
time the Invia Motion
pump can be operated
as usual. A signal will
occur after 72 hours
and the negative pres-
sure will not be maintai-
ned.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

087.4022087.4024

Tabla de contenido