8.
Troubleshooting
Only personnel having received the necessary instruction or training by Invacare
service and maintenance on Jasmine
Symptom
The mast seems loose
Castors are noisy
Hinge between jib and mast
is noisy during operation
The motor is not running
Motor noise, but no
movement in the piston rod
The piston rod only moves
in and not out
The jib motor stops during a
patient transfer
Contact your dealer if the above does not solve your problems.
Lubricate with medically clean oil, e.g. Kemitura Kem Lub KEM-WO 50, order no. 813239.
.
TM
Possible cause
The mast screws
has become
loose
Dust or dirt in
the castors
Missing
lubrication
Hand control
or motor is not
connected
No power on
the battery
The red
emergency stop
is activated
Motor is
damaged
Motor is
damaged
Max. load is
exceeded
18
Remedy
Tighten the screw
Clean the castors
Lubricate the hinge
Insert the plugs fully
Charge or replace the battery
Turn the red button clockwise to
deactivate
Replace the motor
Replace the motor
Reduce the load (and the hoist will
function normally)