TROUBLESHOOTING

SYMPTOMS
T
R
Stand up lift feels loose.
O
U
B
L
E
S
Casters/Brakes noisy or stiff.
H
O
O
Noisy or dry sound from pivots.
T
I
Electric actuator fails to lift when
N
button is pressed.
G
Lift arms will not lower in uppermost
position.
NOTE: If problems are not remedied by the suggested means, please contact your Dealer or Invacare.
TROUBLESHOOTING
FAULTS
Mast/Base joint loose.
Tie - Rods are loose.
Fluff or debris in bearings.
Needs lubrication.
Hand-control or actuator con-
nector loose.
RED emergency stop button
pressed IN.
Batteries low.
Battery not connected properly to
control box.
Electric actuator in need of ser-
vice.
Lift arms require a minimum
weight load to lower from the
uppermost position.
16
SOLUTION
Tighten the bolt, washer and locknut that
secure the mast to the base.
Refer to BASE ADJUSTMENT in the
CARE AND MAINTENANCE section of
this manual.
Refer to CASTERS in the CARE AND
MAINTENANCE section of this manual.
Refer to LUBRICATION in the CARE AND
MAINTENANCE section of this manual.
Check connections.
Rotate RED emergency stop button
CLOCKWISE until it pops out.
Charge batteries. Refer to CHARGING
BATTERIES in the OPERATION section
of this manual.
Reconnect the battery to the control box.
Refer to CHARGING BATTERIES in the
OPERATION section of this manual.
Refer to ELECTRIC ACTUATOR
REPLACEMENT in the CARE AND
MAINTENANCE section of this manual.
Contact your Dealer.
Pull down slightly on the lift arms.

Publicidad

Tabla de contenido

Solución de problemas