Functional Description; Use; Method 1: User Reference; Method 2: Earth Reference - Metrix BioTest Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

4.

FUNCTIONAL DESCRIPTION

4.1. Use

There are two methods used to measure electric fields with the
VX003 / VX0100 testers.

4.1.1. Method 1: user reference

Hold the electric field tester
keeping one of your fingers in
direct contact with the correct
area of the rear face of the
appliance.
To carry out a measurement
refer to the section that
covers your tester.

4.1.2. Method 2: earth reference

Check the presence of earth
using
Connect the electric field
tester to the earth terminal of
an electric socket using the 5
m cable shipped with the
appliance.
To
measurement please refer to
the section covering your
tester.

4.2. Powering up (press "ON")

4.2.1. VX0003

Phase 1: LED test
1.
The green "RANGE" LED and the red "HOLD" LED, as
well as the 2000 V/m LED light up for about 2 secs.
2.
Then all the LEDS light for about 3 secs.
Phase 2: Automatic range configuration at start-up
1.
The green "RANGE" LED becomes red.
2.
Indication of the range: 2000 V/m by default, red scale.
3.
The "HOLD" LED stops.
If the measured values are > 120 % of their nominal
range, then the "RANGE" LED blinks and the
sound signal stops
.

4.2.2. VX0100

1.
"ON" is displayed on the LCD for about. 2 secs.
2.
All the segments are displayed for about. 3 secs.
3.
The measurements are displayed.
If you keep the "ON" key pressed down the
VX0100 displays the software version before
passing on to phase 3.
If the measured values are > 120 % of their nominal
range the VX0100 displays "OL" and the V/m
segment will blink.
28
Chapter II
the
socket
tester.
carry
out
the
VX0003 / VX0100

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Vx0003

Tabla de contenido