23
Table 1: Guidance and manufacturer's declaration - electromagnetic
emissions
The device is intended for use in the electromagnetic environment specified below.
The user of this device should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Table 2: Guidance and manufacturer's declaration - electromagnetic
immunity
The device is intended for use in the electromagnetic environment specified below.
The user of this device should assure that it is used in such an environment.
Immunity test
Electrostatic
±6 kV contact
discharge
±8 kV air
(ESD)
IEC 61000-4-2
Electrical fast
±2 kV for
transient/burst
power supply lines
IEC61000-4-4
±1 kV for
input/output lines
Compliance
Group 1
Class B
N/A
N/A
IEC 60601 test level
ENGLISH
Electromagnetic environment
This device uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
This device is suitable for use in all
establishments. It uses internal power supply
and has no connection with power supply
network.
Compliance level
±6 kV contact
±8 kV air
N/A
Electromagnetic
environment
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
N/A