3. Use the
or
to select
Option
Description
Automatic A consecutive number will be automatically assigned to every
sample.
Manual
A keyboard or a barcode scanner is required to enter the sample
ID name before taking a measurement (maximum 15 characters).
Select the data output
Data can be stored or transferred to a printer or to a PC.
1. From the main menu use the
2. Use the
or
to select Data Output and confirm.
3. Use the
or
to select
Option
Description
Deactivated
Select Deactivated if no printer or PC is connected.
For Printer
Select Dot matrix printer or Thermal printer.
For Computer Select Terminal, ComLabo or ComLabo Easy. The ComLabo
Software controls several modules, pH and conductivity
meters, automatic burettes, Samplers and so on from a
computer. The ComLabo Easy software gets pH and
conductivity data from a PC.
Change the date and time
The date and time can be changed from the Date / Time menu.
1. From the main menu use the
2. Use the
or
to select Date / Time and confirm.
3. Use the and to advance to the next step and use the
change a value. Confirm.
The current date and time will be shown on the display.
10 English
or
to select SYSTEM. Confirm.
or
to select SYSTEM. Confirm.
or

Adjust the display contrast

1. From the main menu use the
2. Use the
or
to select Display contrast and confirm.
3. Use the and to adjust the contrast of the display and confirm.

Adjust the temperature

The temperature measurement can be adjusted at 25 °C (77 °F) and/or
85 °C (185 °F) to increase accuracy.
1. Put the probe and a reference thermometer in a container of water at
approximately 25 °C and allow the temperature to stabilize.
2. Compare the temperature read by the meter with that of the reference
thermometer. The difference is the adjustment value for the meter.
Example: reference thermometer: 24.5 °C; meter: 24.3 °C. Adjustment
value: 0.2 °C.
3. Enter the adjustment value for the 25 °C reading:
a. From the main menu use the
b. Use the
or
c. Use the
or
d. Use the arrow keys to enter the adjustment value for 25 °C.
Confirm.
4. Put the probe and a reference thermometer in a container of water at
approximately 85 °C and allow the temperature to stabilize.
5. Compare the temperature from the meter with that of the reference
thermometer. The difference is the adjustment value for the meter.
a. Use the
or
b. Use the arrow keys to enter the adjustment value for 85 °C.
Confirm.
to
c. Select Save changes and confirm.
Change the stirring settings
The magnetic stirrer can be turned on and the stirring speed can be
changed in the Stirring menu.
or
to select SYSTEM. Confirm.
or
to select SYSTEM. Confirm.
to select Readjust temp. and confirm.
to select 25 °C and confirm.
to select 85 °C and confirm.