9.8.1 Current measured values for ClO
and pH/ORP
The measured values in this menu are only
displayed, if "measurement" has been enabled.
If the AQC-D6 measuring cell has been selected,
Note
Note
only the ClO
measured value is displayed.
2
If the AQC-D11 measuring cell has been selected,
the ClO
and the pH/ORP measured values are
2
displayed.
1. "main menu > service" > [OK].
2. "measurement" > [OK].
measurement
ClO2
(1)
temperature
(2)
pH/ORP
(1)
Only visible if measuring cell AQC-D11 is selected.
(2)
Select pH or ORP if desired, not both
ClO
measured value
2
1. "ClO2" > [OK].
measurement ClO2
CalData-LogBook
measured value
2. Press [Down] to select "measured value", confirm with [OK].
The current ClO
measured value is displayed:
2
– ClO
concentration [mg/l],
2
– current from measuring cell [µA]),
– set measuring range [mg/l)].
measurement ClO2
0.2 mg/l
4.061 µA
0.2 - 1.0 mg/l
Calibration logbook for ClO
2
1. Press [Up] to select "CalData-LogBook", confirm with [OK].
CalData-LogBook
number
1
date
2013-05-24
time
09:01
slope:
-54.2
asym.
11.31
buffer 1
4.01
buffer 2
7.00
Cal temp.
25.0
2. "ClO2" > [OK]. The last 10 calibration data records are listed
in chronological order in "CalData-LogBook". Data record 1 is
the one that was last saved.
3. Press [OK] to display the previous data record.
36
, water temperature
pH value
2
1. "main menu > service" > [OK].
2. "measurement" > [OK].
measurement
ClO2
temperature
pH
(1)
3. pH > [OK].
measurement pH
CalData-LogBook
measured value
4. Press [Down] to select "measured value", confirm with [OK].
The current pH measured value is displayed:
measurement pH
7.20 pH
-30 mV
0.00 - 14.00 pH
Press [Esc] to return to the previous menu.
Calibration logbook for pH
1. Press [Up] to select "CalData-LogBook", confirm with [OK].
CalData-LogBook
number
date
time
slope:
asym.
buffer 1
buffer 2
Cal temp.
2. The last 10 calibration data records are listed in chronological
3. Press [OK] to display the previous data record. Press [Up] or
(1)
Only visible if measuring cell AQC-D11 is selected.
– pH measured value,
– voltage [mV],
– set measuring range [pH].
1
2013-05-24
09:01
-54.2
11.31
4.01
7.00
25.0
order. Data record 1 is the one that was last saved.
[Down] to display all lines with calibration data.