Msa Link Operation; Function Tests On The Instrument; Calibration Check - MSA ALTAIR 5 Manual De Funcionamiento

Ocultar thumbs Ver también para ALTAIR 5:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Setting Backlight
1. Press the
button to enter Setup.
2. Change option by pressing the ▼ or ▲ button.
3. Press the
button to enter.
4. Change timeout by pressing the ▼ or ▲ button.
5. Press the
button to confirm timeout.
Back To Main Menu
There are three options at this point:
the ▼ button
to go to the Sensor Options menu
the ▲ button
to go to the Previous Setup page in the
Instrument Options menu
the
button
to go to the Instrument Options menu.

3.5. MSA Link Operation

Connecting Instrument to PC
1. Switch ON the ALTAIR 5 instrument and align the Datalink
Communication port on the ALTAIR 5 to the IR interface of the PC.
2. Start the MSA Link software on the PC and start the connection by
clicking the connect icon.
See MSA Link documentation for detailed instructions.

3.6. Function Tests on the Instrument

Alarm Test
1. Turn ON the instrument. The user should verify that:
• alarm LEDs flash
• horn sounds briefly
• vibrating alarm triggers briefly.

3.7. Calibration Check

The calibration check is simple and should only take about one minute.
Perform this calibration check before each day's use for each
installed sensor.
3-14